[Metash] TOC-3000 Total Organic Carbon Analyzer

텍스트를 입력해주세요

"A premier wet chemical UV oxidation TOC analyzer designed for high-sensitivity trace organic carbon analysis in pure water and regulatory applications."

The Metash TOC-3000 is an advanced Total Organic Carbon (TOC) analyzer that utilizes wet chemical oxidation combined with high-intensity Ultraviolet (UV) radiation. Managed via intuitive PC software, it is optimized for tracking low-level organic contaminants in drinking water, pharmaceuticals, and environmental applications.

 

 

Key Features
 
1. High-Efficiency UV & Wet Chemical Oxidation
  • Dual Digestion Mechanism: Combines high-intensity UV light with strong chemical oxidants (such as persulfate) to thoroughly break down complex organic compounds without needing high-temperature combustion furnaces.

  • Optimized for Pure Water: Highly sensitive approach that minimizes background system noise, making it perfect for validating trace impurities in ultra-pure matrices.

2. High-Performance Signal and NDIR Detection
  • Custom NDIR Tech: Equipped with a proprietary Non-Dispersive Infrared (NDIR) detector paired with a premium German-manufactured light source for exceptional long-term stability and response linearity.

  • 24-Bit Data Resolution: Utilizes an upgraded 24-bit data acquisition system that significantly broadens the monitoring dynamic range, paired with a fast 32-bit core processing unit.

3. Advanced Gas and Fluid Management
  • 3-Stage Dehydration: Features a multi-stage cooling and drying module to completely strip water vapor out of the carrier gas stream, protecting the internal NDIR cell from moisture degradation.

  • Precise Flow Control: Built-in electronic flow monitoring suppresses carrier gas fluctuations, ensuring highly accurate and reproducible peak integration.

  • Auto-Leak Diagnostics: Automates system pressure tests to instantly flag any micro-leaks, improving experimental safety and data validity.

4. Comprehensive Parameter Testing
  • Multi-Mode Operation: Measures Total Carbon (TC), Total Inorganic Carbon (TIC), Total Organic Carbon (TOC = TC - TIC), and Non-Purgeable Organic Carbon (NPOC).

  • Autosampler Compatibility: Can be paired with an optional multi-position liquid autosampler for seamless, high-throughput batch automation.

 

 

Specifications
 
Analytical Performance
  • Oxidation Method: Wet Chemical Oxidation by UV

  • Detector Type: Customized Non-Dispersive Infrared (NDIR) Detector

  • Measurement Range: 0 to 10,000 mg/L (ppm)

  • Detection Limit: 5 $\mu$g/L (ppb) (Ultra-trace detection limit)

  • System Repeatability: $\le$ 3%

  • Maximum Matrix Salinity: Up to 85 g/L

Gas & Hardware System
  • Carrier Gas Requirement: Nitrogen ($\ge$ 99.995%)

  • Operation Mode: External PC Software Control

  • Power Parameters: AC 220V, 50Hz / AC 110V, 60Hz (Max consumption: 200W)

  • Physical Dimensions: 430 mm W $\times$ 455 mm D $\times$ 440 mm H

  • Net Weight: 25 kg

Applications
  • Pharmaceutical Validation: Validating Purified Water (PW), Water for Injection (WFI), and cleaning validation procedures to ensure compliance with strict international pharmacopeia limits.

  • Drinking & Tap Water Quality: Monitoring disinfection byproduct precursors and organic carbon load changes across tap water production facilities.

  • Semiconductor Cleanrooms: Testing incoming feed water purity and monitoring ultra-pure water (UPW) systems used during microelectronics wafer fabrication.

  • Biomedical Engineering: Routine analysis of raw cell culture fluids and organic loads inside sensitive biotechnology processing setups.